Canon R-F10 User Manual Page 19

  • Download
  • Add to my manuals
  • Print
  • Page
    / 60
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 18
19
6. Control Digital Refractor
and Chart Projector by Keyboard and
Touch Panel
6.1 Controlled by Keyboard
(1) Turn on the power (Digital Refractor, chart projector and keyboard)
(2) Two kinds of insert modes:
Press Print when Auto refractometer finishes test, then press the key AR on
the keyboard, so all information will be automatically transferred to the Digital
Refractor. It will also automatically turn to the Spherical (right eye) and start
the test..
Insert information manually(have no computer in connection), first choose
one among Spherical, cylinder and axis, press it directly. Turn around the big
knob( right:+, left:-) to adjust the degree, or you can also use + and – to adjust
the degree.
(3) Look at the following picture.
For example: CLK Dial means Auxiliary Lenses are in work.
5D ON:
5D: Axis, it turns 5 degrees when dial the knob once; 1D: it will be 1 degree
a time. Both can be set by Auxiliary Lenses.
ON: Chart projector in connection
OFF: Chart projector off
ERR: Chart projector out of connection
Whenever C
0.05
appears on the touch panel, touch Help, the details appears
(when other charts appears, the Help is not in function).
(4) P.D test: Press P.D to start the test, press it again to return when finish the
test